NEWTOM – GIANO HR

Perfect. Vision Unlimited diagnostic potential

  • Powerful diagnosis
  • Newtom HR Technology
  • Patient safety
  • 10″ Touch Console
Enquiry

GiANO HR is the versatile and upgradeable NewTom device, suitable for all radiological needs. Equipped with a complete range of 2D and 3D examinations for dentistry, it offers multiple volumetric examinations specific for maxillofacial surgery, otolaryngology and the study of the cervical area. Always the best image quality thanks to NewTom technology and experience.

Technical data: GiANO HR
Generator type Constant high frequency potential:100180 kHz
Anode voltage 2D: 60 kV – 85 kV | 3D: 90 kV (Pulsed mode)
Anode current 2 mA – 16 mA
Focal spot 0.5 mm (IEC 60336) – Fixed anode
Exposure Control Auto-Adaptative with intensity modulation during rotation –SafeBeamTM Technology
Maximum continuous anode input power 42 W (1:20 at 85 kV/10 mA)
Inherent filtration 2D: >2.5 mm Al eq. (at 85kV) |  3D: 6.5 mm Al eq. (at 90 kV)
Image Acquisition
Detector type 2D: traditional with scintillator (CsI) or Direct Conversion (DCIII technology)

3D: high resolution Amorphous Silicon (CsI)

Image Dynamic Range 2D Standard: 14 bit (16384 grey levels) 2D DC: 16 bit (65536 grey levels) | 3D: 16 bit (65536 grey levels)
Ergonomics
Patient alignment Supported by 4 laser guide lights marking reference planes and height of the FOV
Patient positioning 7 head contact points
Adjustments On-board keypad and/or virtual console for iPad (2-speed height drive)
Examination selection Virtual console on PC, Windows tablet and/or iPad and from Full-Touch 10” on-board console
Notes Easy access for patients in wheelchairs
Connectivity
Connections LAN / Ethernet
Software NNT (ISDP©10003:2020 compliant in accordance with EN ISO/IEC 17065:2012 certificate number 2019003109-1) with Viewer software, free
Supported protocols DICOM 3.0, TWAIN, VDDS
DICOM nodes IHE compliant (Print; Storage Commitment; WorkList; MPPS; Query/Retrieve)
App iPad Virtual control panel for the device and for the NNT 2D viewer
Installation
Minimum available work space requirement 2D and 3D PAN: 1400 x 1200 mm    2D and 3D CEPH: 1400 x 1790 mm
Package dimensions (L) x (D) x (H) in mm Machine Base: 1515 x 1750 x 670 mm – CEPH application: 1030 x 530 x 360 mm
Weight 2D PAN: 155 Kg – 342 lbs |  2D CEPH: 175 Kg – 386 lbs

3D PAN: 155 Kg – 342 lbs |  3D CEPH: 175 Kg – 386 lbs

Accessories Wall bracket even at 45° or floor support, free standing base available User-friendly for patients on wheelchair
Power supply
Voltage | Frequency 115 – 240 Vac, +/- 10% | 50/60 Hz +/- 2 Hz
Maximum absorbed surge current 20 A at 115 V; 12 A at 240 V
Absorbed power in stand-by mode 20 Watt
Notes Automatic adaptation for voltage and frequency
Generator type Constant high frequency potential:100180 kHz
Anode voltage 2D: 60 kV – 85 kV | 3D: 90 kV (Pulsed mode)

1: optional

2: included for versions distributed in the USA and CANADA

3: Specifications subject to change without prior notice. always